| System Bus |
USB 2.0 |
| Bus Information |
Baud Rate, Max. |
384 bps |
Bus Standards Compliance |
USB 1.1 USB 2.0 | System Bus Details |
- USB Type B (4-pin, square) connector
- Plug-and-play USB 2.0, draws +5V power from the USB cable
| | Designed to Replace |
UWDAS TC |
| Mounting Configuration |
Desktop / benchtop External module |
| Direct Sensor Input |
Thermocouple |
| On-board Microprocessor |
Yes |
| On-Board Memory for Processor |
1K |
|
| Number of Channels |
8 |
| Differential Channels |
8 |
| A/D Converter Resolution |
24-bit |
| Max. A/D Speed (throughput) |
384Hz |
| Typ. A/D Speed (throughput) |
384Hz |
| A/D Speed/Bandwidth Details |
Each two channels share a dual-channel 24-bit A/D converter. Each channel pair is sampled twice per second, regardless of the number of channels in use. If all 8 channels are used, then the total throughput is is 16 samples per second, or 384 bps. |
| Cold Junction Compensation |
CJC on all channels |
| Accuracy |
Relative Accuracy |
Typical errors range from +/-0.245 deg. C to +/-0.650deg. C, depending on thermocouple type. Maximum errors range from +/- 0.639 deg. C to +/-2.653 deg. C, depending on thermocouple type. |
| Noise |
CMRR (Common Mode Rejection Ratio) |
100dB min. @ 50/60Hz |
NMRR (Normal Mode Rejection Ratio) |
90dB min. @60Hz | | Input Impedance |
5000MΩ |
| Input Impedance Details |
5 Gigohm min. |
| A/D Special Features |
- Four dual 24-bit Sigma-Delta type A/D converters
- Two CJC high-resolution temperature sensors measure the average temperature at the terminal blocks so that the cold junction voltage can be calculated and the thermocouple reading corrected via software
- CJC Sensor Accuracy: ±25°C typ. & ±0.5°C max. @ 15 to 35°C
-1.0 to +0.5°C max. @ 0 to 70°C
- Open thermocouple detection (OTD)
- 500VDC minimum isolation between field wiring and USB interface
|
|
| Bipolar Input Ranges |
-80mV to +80mV |
| Thermocouple Support |
Thermocouple Inputs |
J-type thermocouple K-type thermocouple T-type thermocouple E-type thermocouple R-type thermocouple S-type thermocouple B-type thermocouple N-type thermocouple |
J-type Range Supported |
-346°F....2192°F
(-210°C....1200°C)
| K-type Range Supported |
-346°F....2501.6°F
(-210°C....1372°C)
| T-type Range Supported |
-328°F....1112°F
(-200°C....600°C)
| E-type Range Supported |
-328°F....1832°F
(-200°C....1000°C)
| R-type Range Supported |
-58°F....3214.58°F
(-50°C....1768.1°C)
| S-type Range Supported |
-58°F....3214.58°F
(-50°C....1768.1°C)
| B-type Range Supported |
482°F....3308°F
(250°C....1820°C)
| N-type Range Supported |
-328°F....2372°F
(-200°C....1300°C)
| | A/D Range Selection |
Fixed |
| Input Range Details |
Range is automatically selected by configuring sensor type in software. |
| Gains |
1 |
| Isolation, Field-to-Logic |
500VDC |
| Isolation Details |
500VDC min. input isolation between field wiring and USB interface |
|
| Triggering Features |
Triggering is fixed at 2 (24-bit) samples per second on each channel. |
|
| Number of I/O Lines |
8 |
| Digital I/O Details |
TTL-compatible, individually configurable as input or output |
| Primary Digital I/O |
Configuration |
8 bidirectional |
Input/Output Selection |
Software | Signal Connector |
Screw terminals | Inputs |
Input Voltage - High (Vih, DC) |
2V....5.5V | Input Voltage - Low (Vil) |
-0.5V....0.8V | Digital Input Features |
Transfer rate is software paced, typically 50 port reads or single-bit reads per second. | Outputs |
Output Voltage - High (Voh) |
3.8V....5V | Output Voltage - Low (Vol) |
0mV....0.7V | Output Current Sink, max. |
2.5mA | Output Current Sink Details |
2.5mA @ 0.7V | Output Current Source Details |
-2.5mA @ 3.8V | Output/Relay Details |
Transfer rate is software paced, typically 100 port writes or single-bit writes per second. Power-on reset is input mode. | |
| Software Included Free with Product |
calibration software |
| Software Included - Features |
InstaCal calibraton software included. Other software under development. |
| Operating System Compatibility |
Windows XP Windows 2000 Windows Me Windows 98SE |
| Software Compatibility |
Universal Driver Library DASYLab® LabVIEW® |
|
| Connector #1 |
Quantity |
52 |
Type |
Screw terminals | Location |
On module | | Connector & Cabling Details |
- Includes a 3-meter (max.) USB cable
- 52 screw terminals arranged in a block of 10 and of 16 screw terminals on
either side of the module, with a CJC sensor on each
side, between each set of blocks
- Wire Gauge Range: 16 to 30 AWG
|
|
| Construction Material |
Industrial plastic |
|
| Power Consumption (typical) |
70mA max. |
| +5V |
70mA |
| Power Supply Notes |
- USB bus; draws +5V power from the USB cable
- Two +5V terminals (output) are isolated (500VDC) from the USB +5V
power
|
|
| Other Control Features |
Power and data LED indicator |
|
| Operating Temperature |
32°F....158°F
(0°C....70°C)
|
| Storage Temperature |
-40°F....185°F
(-40°C....85°C)
|
| Relative Humidity, noncondensing |
0%....90% |
|
| Safety/Approvals |
Meets CE standards |
|
| Width |
3.5"
(88.9mm)
|
| Height |
1.4"
(35.56mm)
|
| Depth |
5"
(127mm)
|